Common Problems and Solutions for Dual 85 Testing of Temperature and Humidity Test Chambers

Common Problems and Solutions for Dual 85 Testing of Temperature and Humidity Test Chambers

December 09, 2025

    This guide summarizes key issues and solutions for temperature & humidity test chambers during dual 85 testing (85℃, 85% RH), focusing on quick troubleshooting and long-term maintenance for operators and customers.

I. Core Issues & Resolution System

Each issue includes root cause, immediate solution, and long-term prevention (except for defect exposure).

1. Failure to Reach 85% RH

Root Cause: Compressor over-dehumidifies—evaporator temperature below dew point (79-80℃) causes condensation, outpacing humidification.

Immediate Solution: Disable compressor; run only heating and humidification to maintain 85℃ and avoid condensation.

Prevention: Regularly calibrate humidifier atomization volume and check evaporator temperature sensor accuracy.

2. Temperature Out-of-Control/Fluctuations

Root Cause: Mainly faulty solid-state relays (keeps heating elements on), plus aging heating tubes or defective temperature sensors.

Immediate Solution: Cut power; inspect circuit with multimeter (relay on-off, heating tube damage, sensor accuracy) and replace faulty parts.

Prevention: 10-15min pre-calibration before testing; replace vulnerable parts (relays, sensors) every 1000 operating hours.

3. Water Marks/Condensation on Products

Root Cause: Residual oil/dust on products, or dense samples blocking air ducts (poor circulation, localized high humidity).

Immediate Solution: Pause test; clean products with anhydrous ethanol, dry, and reposition samples for proper spacing.

Prevention: Establish pre-test cleaning standards; clean air duct filters regularly and optimize sample rack layout.

4. Incorrect Operation Logic

Root Cause: Operator errors—misactivated refrigeration, wrong parameters (fast heating, delayed humidification), or mismatched operation modes.

Immediate Solution: Reset program (disable refrigeration, set 85℃/85% RH, sync humidification with heating).

Prevention: Develop SOPs; new employees must pass operation assessments (including simulated faults).

5. Material/Process Defect Exposure

Root Cause: Extreme environment accelerates aging—e.g., poor heat-resistant EVA film hydrolyzes (yellowing, delamination); leaky electronic packages cause corrosion/short circuits.

Measures: Inspect products post-test, record defects; judge qualification by material standards; optimize processes (e.g., use hydrolysis-resistant EVA, enhance sealant density).